CR Phantom

The CR Phantom can test all the important performance characteristics of a CR scanner system including basic spatial resolution (unsharpness), contrast, MTF, laser beam jitter, scanner slipping and scanner shading. These tests are described in detail in the ISO 16371-1 and ASTM E 2445 standards and should be performed periodically to ensure proper and accurate system operation. For easy use of the CR Phantom, a test procedure and the D-Tect Viewer software are available for free download.

To meet the requirements of these standards, the DÜRR NDT CR Phantom includes two duplex wire- type IQIs and measuring points for shading correction positioned in both axis directions (horizontal and vertical). This allows all the required information to be mapped on the imaging plate with a single X-ray exposure – the CR Phantom does not need to be rotated to obtain the data of the second axis, resulting in more accurate test scores and significant time savings.

If you are interested in this product, please contact us!